80 OUTPUT 718;"AMPCOR?;"<br />

90 ENTER 718;A$<br />

100 PRINT A$<br />

110' END<br />

Description<br />

AMPCOR Amplitude Correction<br />

Because AMPCOR is ofi “0,O” is displayed.<br />

Use AMPCOR to compensate for frequency-dependent amplitude variations at the spectrum<br />

analyzer input. Up to 79 pairs of frequency-amplitude correction points can be entered. The<br />

frequency values entered must either be equal or in increasing order, or an error condition<br />

results. Whenever AMPCOR is on, the correction values are added to all measurement<br />

results. Executing “AMPCOR ON;” or entering frequency <strong>and</strong> amplitude corrections, turns<br />

on the amplitude correction factors. Performing an instrument preset (IP) or turning off<br />

the spectrum analyzer sets AMPCOR to OFF. (Setting AMPCOR to OFF does not change the<br />

frequency-amplitude correction factors that have been entered.)<br />

The values of the correction points are applied across the active measurement range. Between<br />

points, the correction values are interpolated. When measuring at frequencies outside the<br />

first <strong>and</strong> last correction points, these values are used as the correction value. If you do not<br />

want any amplitude correction outside of the first <strong>and</strong> last correction points, set the amplitude<br />

correction to 0 at the frequencies that are outside of the first <strong>and</strong> last correction values.<br />

With the Analog + display mode: When the Analog+ display mode is turned on, AMPCOR<br />

applies only one amplitude correction across the displayed frequency span. When the Analog+<br />

display mode is turned on, the amplitude correction for the current center frequency is used<br />

over the entire frequency span.<br />

Amplitude correction factors can be stored in spectrum analyzer memory with the SAVET or<br />

SAVRCLN comm<strong>and</strong>s, or on the memory card with the STOR or SAVRCLN comm<strong>and</strong>s. The<br />

amplitude correction factors can be edited <strong>and</strong> viewed with the Amp Cor softkey functions.<br />

Query Response<br />

AMPCOR? returns the frequency <strong>and</strong> amplitude correction pairs.<br />

frequency amp I i tude<br />

/ \<br />

In-<br />

Returned values are 0,O when AMPCOR is set to OFF<br />

Programming Comm<strong>and</strong>s 5-63