OP 67<br />

Title<br />

DOSIMETRy OF THREE CbCT DEVICES uSING RADIOCHROMIC FILM<br />

Authors<br />

A. AL-OKSHI 1 , M. NILSSON 1 , A. PETERSSON 1,2 , M. WIESE 2 , C. LINDH 1<br />

Affiliations<br />

1 Departnet of Oral and Maxillofacial Radiology, Faculty of Odontology, Malmö University,<br />

Malmö, SWEDEN, 2 Department of Oral Radiology, School of Dentistry, University of<br />

Copenhagen, Copenhagen, DENMARK<br />

Body<br />

Objective Cone Beam Computed Tomography (CBCT) is being used with incre<strong>as</strong>ing frequency in<br />

the practice of Oral and Maxillofacial Radiology. The radiation doses to patients vary between<br />

machines depending on differences in exposure parameters and t<strong>here</strong> is a wide range of effective<br />

doses reported for CBCT devices. To provide comparative me<strong>as</strong>urements of the effective<br />

dose from the CBCT devices Veraviewpocs 3De, ProMax 3D and NewTom VGi, and to compare<br />

the doses from examinations of three clinical situations by the use of radiochromic film <strong>as</strong><br />

dosimeter.<br />

Methods Radiochromic films (Gafchromic® XR-QA 2) were placed between selected layers of<br />

the head and neck of a tissue-equivalent human skull (RANDO phantom). Depending on availability,<br />

medium and smaller FOV scanning modes were used. Effective dose w<strong>as</strong> estimated using<br />

the 2007 ICRP formalism. Exposure parameters were set by the automatic exposure control<br />

function of the units.<br />

Results The lowest effective dose w<strong>as</strong> observed for ProMax 3D 5x4 cm ( 9 µSv),the highest for<br />

the NewTom VGi 8x8 cm – high resolution (118 µSv).<br />

Veraviewepocs 3 De<br />

4x4 cm Anterior Maxilla Standard 20µSV<br />

4x4 cm Posterior Mandible Standard 19µSV<br />

ProMax 3D<br />

5x4 cm Anterior Mandible Standard 9µSV<br />

NewTom VGi<br />

8x8 cm TMJ – Standard 43µSV<br />

8x8 cm TMJ – High 118µSV<br />

8x12cm TMJ – Zoom mode 48µSV<br />

Conclusions CBCT dose varies substantially depending on the device, FOV, and selected resolution<br />

factor. This study also demonstrates the fe<strong>as</strong>ibility of using radiochromic films for dental<br />

CBCT dosimetry.<br />

Keywords<br />

Cone Beam CT, Radiation dosage, Film dosimetry<br />