Self tests (SST and EST)<br />

No sound Confirm that you do not hear alarm by pressing<br />

ACCEPT.<br />

Is backup alarm<br />

ON?<br />

<strong>Table</strong> 4-8: Prompts during EST (continued)<br />

Prompt Your response<br />

Confirm that you hear alarm by pressing ACCEPT.<br />

7. When the last test in the sequence is complete, the EST<br />

completion status is displayed (<strong>Table</strong> 4-9). Respond as indicated.<br />

If certain tests do not pass, safe ventilation may be<br />

compromised, and the ventilator is rendered inoperative until it<br />

is repaired. This is called a ventilator failure. If certain other<br />

tests do not pass, but ventilation might not be compromised<br />

(subject to a trained operator’s evaluation), this is called a<br />

ventilator fault.<br />

NOTE:<br />

Faults detected during EST may not require the ventilator to<br />

be serviced or removed from use immediately. A trained<br />

operator, however, must evaluate the situation and determine<br />

whether the ventilator can be used. Failures detected during<br />

EST require immediate servicing and removal <strong>of</strong> the ventilator<br />

from clinical use.<br />

4-24 10066984 Rev. D 700 Series Ventilator Operator’s Manual