4.1 Short self test (SST)<br />
<strong>Table</strong> 4-2 summarizes the functions <strong>of</strong> SST.<br />
Self tests (SST and EST)<br />
NOTE:<br />
• <strong>Covidien</strong> recommends that you run SST every 15 days,<br />
between patients, and when you change the ventilator<br />
breathing circuit. <strong>Covidien</strong> recognizes that the protocol for<br />
running SST varies widely among healthcare institutions.<br />
It is not possible for <strong>Covidien</strong> to specify or require specific<br />
practices that will meet all needs, or to be responsible for<br />
the effectiveness <strong>of</strong> those practices.<br />
• When SST asks you to do something (for example, press<br />
ACCEPT or block the patient wye), the ventilator will wait<br />
indefinitely for your response.<br />
Follow these steps to run SST:<br />
1. Turn on the ventilator. If ventilation has already begun, you<br />
must turn <strong>of</strong>f the ventilator, then turn it back on without<br />
starting ventilation. This allows you access to the Self test menu<br />
function.<br />
If you select the Self test function during ventilation the<br />
window displays this message:<br />
N/A in ventilation<br />
2. Press MENU, turn the knob to select Self test, then press ACCEPT.<br />
The usual 30-second timeout for your input is disabled during<br />
the entry to the Self test function.<br />
3. The window displays this message:<br />
Short self test<br />
Turn knob to view<br />
4. Turn the knob to select Short self test (you can also select<br />
Extended self test). The window displays this message:<br />
Short self test<br />
Vent warming 09:59<br />
CLEAR to bypass<br />
700 Series Ventilator Operator’s Manual 10066984 Rev. D 4-3