Alarm testing<br />

17. % O2 LOW alarm test: Select the following settings:<br />

Mode A/C<br />

RESPIRATORY RATE 20/min<br />

TIDAL VOLUME 1000 ml<br />

PEAK FLOW 60 L/min<br />

PLATEAU 0 s<br />

HIGH PRESSURE<br />

% O2 90 cmH2O (90 hPa)<br />

100 %<br />

18.<br />

Allow 36 seconds for % O2 setting to stabilize (the % O2 setting is<br />

stable when the ventilator has delivered 12 L <strong>of</strong> volume).<br />

Disconnect the oxygen inlet supply.<br />

19. The ventilator annunciates a LOW O2 SUPPLY alarm within<br />

2 breaths, and a % O2 LOW alarm within approximately<br />

30 seconds.<br />

20. Reconnect the oxygen inlet supply. The LOW O2 SUPPLY and<br />

% O2 LOW alarms should autoreset within approximately<br />

30 seconds.<br />

The alarm test is complete.<br />

E-4 10066984 Rev. D 700 Series Ventilator Operator’s Manual