31.10.2012 Views

Table of Contents - Covidien

Table of Contents - Covidien

Table of Contents - Covidien

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Getting started<br />

3.2 Selecting ventilator settings<br />

NOTE:<br />

Each keypress should result in a “blip” sound; otherwise, the<br />

main alarm speaker may be faulty. If a keypress does not result<br />

in this sound, turn the ventilator <strong>of</strong>f then on to run POST and<br />

verify the proper functioning <strong>of</strong> the speaker.<br />

1. Select the mode: To select a mode, press A/C, SIMV, or SPONT.<br />

(Appendix D gives a detailed explanation <strong>of</strong> these ventilation<br />

modes.)<br />

A/C<br />

SIMV<br />

SPONT<br />

In assist/control (A/C) mode, you can only select volume<br />

control ventilation (VCV) or pressure control<br />

ventilation (PCV) settings (PCV is only available on the<br />

760 Ventilator).<br />

In synchronous intermittent mandatory ventilation<br />

(SIMV) mode, you can select both VCV or PCV (760 only)<br />

and pressure support ventilation (PSV) settings.<br />

In spontaneous (SPONT) mode, you can only select PSV<br />

settings.<br />

2. Confirm or change settings: For every flashing key light, you<br />

must touch the key (and adjust the setting if necessary) before<br />

you can apply the settings. You can confirm or change settings<br />

in any order.<br />

Once you press a key, the proposed<br />

setting flashes to indicate that the<br />

knob is linked to the selected<br />

setting. The setting’s LCD window<br />

shows the setting and the message<br />

window displays the current<br />

setting, range, units, name <strong>of</strong> the<br />

setting, and associated parameter.<br />

Turn the knob to change the<br />

setting or press another key to<br />

leave the setting as is. (The setting<br />

then stops flashing.)<br />

Press each flashing key<br />

and turn the knob to<br />

adjust.<br />

7-00066<br />

3-4 10066984 Rev. D 700 Series Ventilator Operator’s Manual