When you<br />

see this<br />

message...<br />

OCCLUSION High-priority alarm.<br />

Ventilator breathing circuit or<br />

inspiratory or expiratory filters<br />

occluded. Ventilator detects<br />

above-normal difference<br />

between inspiratory and<br />

expiratory pressure transducers.<br />

The ventilator enters occlusion<br />

cycling mode.<br />

Autoreset when the ventilator<br />

no longer detects an occlusion.<br />

PARTIAL<br />

OCCLUSION<br />

SETUP TIME<br />

ELAPSED<br />

<strong>Table</strong> 7-1: Clinical alarms (continued)<br />

It means... Do this...<br />

High-priority alarm.<br />

Ventilator breathing circuit or<br />

inspiratory or expiratory filters<br />

partially occluded. Ventilator<br />

detects above-normal difference<br />

between inspiratory and<br />

expiratory pressure transducers.<br />

The ventilator continues normal<br />

ventilation.<br />

Autoreset when partial<br />

occlusion no longer exists for<br />

two consecutive breaths.<br />

High-priority alarm.<br />

At least 30 seconds have<br />

elapsed since you pressed a key<br />

or turned the knob (occurs at<br />

power-on only).<br />

Autoreset when you accept<br />

proposed settings.<br />

Alarm handling<br />

Check patient.<br />

Check ventilator<br />

breathing circuit and<br />

inspiratory and<br />

expiratory filters for<br />

occlusions or kinks.<br />

Empty excess water<br />

from tubes.<br />

Press the alarm reset<br />

key.<br />

If this does not resolve<br />

the problem, provide<br />

alternate ventilation<br />

and contact service.<br />

Check the patient.<br />

Check the ventilator<br />

breathing circuit for<br />

water, kinks, or other<br />

blockages.<br />

If this does not resolve<br />

the problem, provide<br />

alternate ventilation<br />

and contact service.<br />

Check the patient. Be<br />

sure to complete<br />

ventilator setup before<br />

connecting the<br />

ventilator breathing<br />

circuit to the patient.<br />

Select appropriate<br />

ventilator settings.<br />