Self tests (SST and EST)<br />
If the ventilator has already been running for more than<br />
10 minutes before you select Short self test, this message is<br />
bypassed (see step 5). If the ventilator was recently running and<br />
is already warmed up, you can press CLEAR to override the<br />
Vent warming message and begin SST. <strong>Covidien</strong> cannot<br />
guarantee the accuracy <strong>of</strong> test results in this case, however.<br />
5. The window displays this message:<br />
Is pt disconnected?<br />
ACCEPT to proceed<br />
6. Press ACCEPT to confirm that no patient is connected to the<br />
ventilator. (Or press CLEAR to back up through the menu.)<br />
The window displays this message:<br />
Block wye<br />
7. Block the patient wye and then press ACCEPT.<br />
8. The ventilator runs power-on self-test (POST) and displays this<br />
message for a few seconds:<br />
POST running...<br />
Warning<br />
If a single beep is not audible during POST, the main audible<br />
alarm may be malfunctioning. Remove the ventilator from use<br />
and contact service.<br />
NOTE:<br />
• If the ventilator is running on ac but the battery is low, upon<br />
power up it may begin ventilating using the last valid settings<br />
and an ABNORMAL RESTART alarm may occur.<br />
• During POST it is recommended that you not switch <strong>of</strong>f<br />
ventilator power.<br />
9. The message window then displays this message:<br />
humidification device type<br />
Choose humidifier<br />
where the humidification device type is the most recently<br />
selected type, as follows:<br />
4-4 10066984 Rev. D 700 Series Ventilator Operator’s Manual