Self tests (SST and EST)<br />

NOTE:<br />

<strong>Covidien</strong> recommends that you always run the full EST before<br />

placing the ventilator into operation following service and as<br />

part <strong>of</strong> the ventilator’s routine performance verification. For<br />

more information on performance verification, see the<br />

700 Series Ventilator System Service Manual.<br />

Follow these steps to run EST:<br />

1. Set up the ventilator as for normal operation, complete with<br />

humidifier, if applicable, and leak-tight ventilator breathing<br />

circuit.<br />

2. Turn on the ventilator. If ventilation has already begun, you<br />

must turn <strong>of</strong>f the ventilator, then turn it back on without<br />

starting ventilation. This allows you access to the Self test menu<br />

function.<br />

3. Press MENU, turn the knob to select Self test, then press ACCEPT.<br />

4. Turn the knob to select Extended self test. The window displays<br />

the messages shown in <strong>Table</strong> 4-5.<br />

<strong>Table</strong> 4-5: EST setup messages<br />

Message Description<br />

Extended self<br />

test Vent<br />

warming xx:xx<br />

CLEAR to bypass<br />

Is pt<br />

disconnected?<br />

This message is displayed (followed by a countdown) if<br />

you just turned on the ventilator. The ventilator must<br />

be on for 10 minutes before running EST so that<br />

component temperatures can stabilize.<br />

If the ventilator was recently running and is already<br />

warmed up, you can press CLEAR to override the Vent<br />

warming message and begin SST. <strong>Covidien</strong> cannot<br />

guarantee the accuracy <strong>of</strong> test results in this case,<br />

however.<br />

Confirm that patient is disconnected by pressing<br />

ACCEPT; or press CLEAR to return to service menu.<br />

Block wye Install no. 2 stopper and then press ACCEPT.<br />

4-14 10066984 Rev. D 700 Series Ventilator Operator’s Manual