4.2 Extended self test (EST)<br />

Self tests (SST and EST)<br />

To run EST, you must have the hardware listed in <strong>Table</strong> 4-4.<br />

Hardware<br />

description<br />

Ventilator<br />

breathing circuit<br />

Stopper, wye<br />

(no. 2)<br />

Stopper,<br />

inspiration port<br />

(no. 3)<br />

Oxygen source,<br />

40 to 90 psi<br />

(275 to 620 kPa)<br />

Warning<br />

<strong>Table</strong> 4-4: EST hardware requirements<br />

Manufacturer/model or <strong>Covidien</strong> part number<br />

To use as test circuit: G-061208-00 or equivalent (adult,<br />

reusable, without heated wire)<br />

NOTE:<br />

To ensure that compliance compensation<br />

functions correctly, the user must run EST or SST<br />

with the circuit configured as intended for use<br />

on the patient.<br />

G-061574-00 or local supplier<br />

G-061575-00 or local supplier<br />

Local supplier<br />

Due to excessive restriction <strong>of</strong> Air Liquide*, Australian, and<br />

Dräger* hose assemblies, reduced FIO 2 levels can result if you use<br />

oxygen inlet pressures < 50 psi (345 kPa). Make sure oxygen inlet<br />

pressure is ≥ 50 psi (345 kPa) when using these hose assemblies, to<br />

maintain correct FIO 2 levels.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 4-13