Maintenance<br />

Figure A-1. Removing/replacing the collector vial<br />

A.2.3 Daily or as required: in-line water traps<br />

Drain as required.<br />

7-00076<br />

A.2.4 As necessary: oxygen sensor calibration<br />

Use the MENU key’s Oxygen sensor function, then select Calibrate<br />

O 2 sensor to perform a two-point calibration <strong>of</strong> the oxygen sensor<br />

anytime you want to verify proper sensor operation. You can only<br />

calibrate the oxygen sensor immediately following powerup, before<br />

a patient is connected. (Chapter 6 tells you how to use the MENU<br />

key.)<br />