Getting started<br />

Ventilation begins according to the displayed settings and the<br />

existing alarm limits. You should review all alarm limits and<br />

change them if necessary. (Section 3.3 tells you how.) You can<br />

change alarm limits or ventilator settings at any time during<br />

normal ventilation.<br />

Warning<br />

In PSV it is impossible to set support pressure + PEEP to<br />

deliberately cause a HIGH PRESSURE alarm. Although <strong>Covidien</strong><br />

does not recommend doing so, it is possible in VCV to set a tidal<br />

volume that causes breath truncation and a HIGH PRESSURE<br />

alarm. To avoid activating a HIGH PRESSURE alarm, select the<br />

appropriate TIDAL VOLUME and HIGH PRESSURE alarm settings.<br />

3.3 Viewing and changing alarm settings<br />

You can view alarm settings at any time without affecting<br />

ventilation by pressing the appropriate key. The ALARM SETTINGS<br />

window displays the value <strong>of</strong> the alarm being viewed or changed.<br />

The alarm setting flashes for up to 30 seconds, and you can change<br />

an alarm setting any time its setting is flashing in the window.<br />

All alarm settings are in effect during ventilation (except for<br />

LOW INSP PRESSURE, which is not active in SPONT mode). You can<br />

change alarm settings at any time during ventilation. You can only<br />

change one alarm setting at a time.<br />

To change a setting: touch, turn, ACCEPT.<br />

1. Touch the key: The key light turns<br />

on, and the current alarm setting<br />

flashes in the window for 30<br />

seconds. The message window<br />

shows the current setting, range,<br />

and units <strong>of</strong> the alarm.<br />

2. Turn the knob to adjust the alarm<br />

setting (this only works when the<br />

alarm setting is flashing in the<br />

window). The ventilator will beep if<br />

you attempt to select a setting<br />

above or below the setting range.<br />

7-00066<br />

7-00068<br />

3-6 10066984 Rev. D 700 Series Ventilator Operator’s Manual