Getting started<br />
Ventilation begins according to the displayed settings and the<br />
existing alarm limits. You should review all alarm limits and<br />
change them if necessary. (Section 3.3 tells you how.) You can<br />
change alarm limits or ventilator settings at any time during<br />
normal ventilation.<br />
Warning<br />
In PSV it is impossible to set support pressure + PEEP to<br />
deliberately cause a HIGH PRESSURE alarm. Although <strong>Covidien</strong><br />
does not recommend doing so, it is possible in VCV to set a tidal<br />
volume that causes breath truncation and a HIGH PRESSURE<br />
alarm. To avoid activating a HIGH PRESSURE alarm, select the<br />
appropriate TIDAL VOLUME and HIGH PRESSURE alarm settings.<br />
3.3 Viewing and changing alarm settings<br />
You can view alarm settings at any time without affecting<br />
ventilation by pressing the appropriate key. The ALARM SETTINGS<br />
window displays the value <strong>of</strong> the alarm being viewed or changed.<br />
The alarm setting flashes for up to 30 seconds, and you can change<br />
an alarm setting any time its setting is flashing in the window.<br />
All alarm settings are in effect during ventilation (except for<br />
LOW INSP PRESSURE, which is not active in SPONT mode). You can<br />
change alarm settings at any time during ventilation. You can only<br />
change one alarm setting at a time.<br />
To change a setting: touch, turn, ACCEPT.<br />
1. Touch the key: The key light turns<br />
on, and the current alarm setting<br />
flashes in the window for 30<br />
seconds. The message window<br />
shows the current setting, range,<br />
and units <strong>of</strong> the alarm.<br />
2. Turn the knob to adjust the alarm<br />
setting (this only works when the<br />
alarm setting is flashing in the<br />
window). The ventilator will beep if<br />
you attempt to select a setting<br />
above or below the setting range.<br />
7-00066<br />
7-00068<br />
3-6 10066984 Rev. D 700 Series Ventilator Operator’s Manual