Specifications<br />

Measuring<br />

and display<br />

devices<br />

(continued)<br />

<strong>Table</strong> C-5: Technical specifications (continued)<br />

Purpose: Inspiratory time (760 Ventilator<br />

only)<br />

Sensing position: Microprocessor<br />

Range: 0.00 to 9.90 s<br />

Type: Inspiratory time measurement<br />

Purpose: Exhaled tidal volume<br />

Sensing position: Exhalation limb<br />

Range: 0 to 9 L<br />

Type: Differential pressure<br />

pneumotach<br />

Purpose: Total minute volume<br />

Sensing position: Exhalation limb<br />

Range: 0 to 99 L<br />

Type: Differential pressure<br />

pneumotach<br />

Purpose: Delivered volume (760 Ventilator<br />

only)<br />

Sensing position: Exhalation limb<br />

Range: 0 to 3000 ml<br />

Type: Differential pressure<br />

pneumotach<br />

Purpose: Spontaneous minute volume<br />

Sensing position: Exhalation limb<br />

Range: 0 to 99 L<br />

Type: Calculated from expiratory flow<br />

measurements<br />

Purpose: FIO2 sensor<br />

Sensing position: Inspiratory manifold<br />

Range: 18 to 109%<br />

Type: Galvanic cell<br />

C-8 10066984 Rev. D 700 Series Ventilator Operator’s Manual