If the EST<br />

result is:<br />

<strong>Table</strong> 4-9: EST completion status<br />

It means: Do this:<br />

EST passed All tests were<br />

performed and all<br />

passed.<br />

EST<br />

incomplete<br />

All tests passed, but<br />

some tests were<br />

skipped. The skipped<br />

tests were passed on a<br />

previous run.<br />

Self tests (SST and EST)<br />

To retest from start <strong>of</strong> test<br />

sequence, press alarm reset key.<br />

To exit EST and resume<br />

ventilation, press ACCEPT.<br />

Unblock the wye as directed, then<br />

press ACCEPT. POST now runs, and<br />

ventilation resumes at last valid<br />

settings.<br />

To retest from the start <strong>of</strong> the<br />

test sequence, press the alarm<br />

reset key.<br />

To exit EST and resume<br />

ventilation, press MANUAL INSP.<br />

You are then asked whether you<br />

want to use the override feature.<br />

Press MANUAL INSP again to<br />

complete the override. Unblock<br />

the wye as directed, then press<br />

ACCEPT. POST now runs, and<br />

ventilation resumes at last valid<br />

settings.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 4-25