Self tests (SST and EST)<br />

<strong>Table</strong> 4-7: Key functions during EST<br />

Key Function<br />

ACCEPT<br />

CLEAR<br />

MANUAL<br />

INSP<br />

7-00424<br />

7-00423<br />

7-00072<br />

To confirm that a requested action is complete.<br />

Repeat a test or return to prompt at start <strong>of</strong> a test.<br />

To override an Incomplete or Fault test result.<br />

To restart from the beginning <strong>of</strong> SST.<br />

To stop testing and skip to the end <strong>of</strong> SST.<br />

NOTE:<br />

If the tubing or humidifier type or ET tube size<br />

that you’ve selected has changed from the<br />

previous time SST was run, this key is temporarily<br />

disabled until the Circuit comp test is complete.<br />

6. Some tests display additional prompts, some <strong>of</strong> which require<br />

your response. These are listed in <strong>Table</strong> 4-8.<br />

<strong>Table</strong> 4-8: Prompts during EST<br />

Prompt Your response<br />

Disconnect O 2<br />

supply<br />

Reconnect O 2<br />

supply<br />

Disconnect ventilator from oxygen supply.<br />

Press ACCEPT.<br />

Reconnect ventilator to oxygen supply. Press ACCEPT.<br />

4-22 10066984 Rev. D 700 Series Ventilator Operator’s Manual