5.5 The 100% O 2 and MANUAL INSP keys<br />

Once ventilation begins<br />

The 100% O 2 key delivers 100% oxygen to the patient for 2 minutes<br />

at any time except during apnea ventilation. Pressing CLEAR at any<br />

time after the 100% O 2 maneuver begins cancels the maneuver and<br />

the oxygen concentration returns to the set % O 2 .<br />

1. Press 100% O2 : The key lights and<br />

delivered oxygen concentration<br />

increases to 100%.<br />

2. Once the oxygen concentration<br />

reaches 100%, a 2-minute timer<br />

starts and the ventilator delivers<br />

100% O2 to the patient. Pressing<br />

the 100% O2 key again restarts the<br />

2-minute timer.<br />

3. When the 2-minute interval has elapsed, the 100% O 2 key light<br />

turns <strong>of</strong>f and the oxygen concentration returns to the set % O 2 .<br />

NOTE:<br />

The displayed EXHALED VOLUME for the first breath after you<br />

press the 100% O 2 key may be higher than the actual volume<br />

exhaled from the patient. The high displayed volume is a<br />

result <strong>of</strong> an initial flush <strong>of</strong> the ventilator with 100% O 2 and<br />

does not reflect actual volume delivered to the patient.<br />

The MANUAL INSP key delivers one<br />

mandatory breath to the patient<br />

according to the current settings (in<br />

A/C or SIMV mode) or the current<br />

apnea settings (in SPONT mode).<br />

You can deliver a MANUAL INSP at<br />

any time during the exhalation<br />

phase <strong>of</strong> a breath. (To deliver a<br />

7-00086<br />

MANUAL INSP, 200 ms <strong>of</strong> the exhalation phase must have<br />

elapsed, and expiratory flow must have decreased to 50% <strong>of</strong><br />

peak inspiratory flow. If you press INSP PAUSE during SPONT<br />

mode, a manual inspiration is delivered when 200 ms <strong>of</strong> the<br />

exhalation phase have elapsed, and expiratory flow is less than<br />

2 L/min for 50 ms) MANUAL INSP is not functional during apnea<br />

ventilation.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 5-9<br />