Introduction<br />
1.3.2 PATIENT DATA<br />
The PATIENT DATA section <strong>of</strong> the keyboard allows you to view the<br />
pressure, breath timing, and volume <strong>of</strong> the patient’s breath. You can<br />
also view the alarm settings. A lighted key indicates that a<br />
measurement is selected, and its value appears in the display<br />
window. Values are continuously displayed and updated during<br />
ventilation.<br />
<strong>Table</strong> 1-4 summarizes the functions <strong>of</strong> the keys and indicators in the<br />
PATIENT DATA section <strong>of</strong> the keyboard.<br />
Key/<br />
indicator<br />
Pressure<br />
MEAN<br />
PRESSURE<br />
PEAK<br />
PRESSURE<br />
PLATEAU<br />
PRESSURE<br />
(760 only)<br />
PEEP/CPAP<br />
(760 only)<br />
<strong>Table</strong> 1-4: 700 Series Ventilator keyboards: PATIENT DATA<br />
Function Range<br />
Shows the calculated value <strong>of</strong><br />
ventilator breathing circuit<br />
pressure over an entire<br />
respiratory cycle. Updated at the<br />
beginning <strong>of</strong> each breath.<br />
Shows the maximum pressure<br />
measured during inspiration.<br />
Updated at the beginning <strong>of</strong><br />
each expiratory phase. (Default<br />
pressure display.)<br />
Shows the pressure measured at<br />
the end <strong>of</strong> the plateau period <strong>of</strong><br />
a mandatory inspiration<br />
(whether the inspiration is in a<br />
regular VCV breath or is part <strong>of</strong><br />
an inspiratory pause maneuver).<br />
Updated at the beginning <strong>of</strong><br />
each expiratory phase. The<br />
PRESSURE display shows a blank<br />
if the ventilator does not detect<br />
a stable plateau pressure.<br />
Shows the pressure measured at<br />
the expiratory limb before any<br />
inspiratory effort. Updated at<br />
the beginning <strong>of</strong> each<br />
inspiratory phase.<br />
0 to 99 cmH 2O<br />
(0 to 99 hPa)<br />
Accuracy:<br />
± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />
0 to 140 cmH 2O<br />
(0 to 140 hPa)<br />
Accuracy:<br />
± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />
0 to 140 cmH 2O<br />
(0 to 140 hPa)<br />
Accuracy:<br />
± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />
0 to 140 cmH 2O<br />
(0 to 140 hPa)<br />
Accuracy:<br />
± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />
1�24 10066984 Rev. D 700 Series Ventilator Operator’s Manual