Introduction<br />

1.3.2 PATIENT DATA<br />

The PATIENT DATA section <strong>of</strong> the keyboard allows you to view the<br />

pressure, breath timing, and volume <strong>of</strong> the patient’s breath. You can<br />

also view the alarm settings. A lighted key indicates that a<br />

measurement is selected, and its value appears in the display<br />

window. Values are continuously displayed and updated during<br />

ventilation.<br />

<strong>Table</strong> 1-4 summarizes the functions <strong>of</strong> the keys and indicators in the<br />

PATIENT DATA section <strong>of</strong> the keyboard.<br />

Key/<br />

indicator<br />

Pressure<br />

MEAN<br />

PRESSURE<br />

PEAK<br />

PRESSURE<br />

PLATEAU<br />

PRESSURE<br />

(760 only)<br />

PEEP/CPAP<br />

(760 only)<br />

<strong>Table</strong> 1-4: 700 Series Ventilator keyboards: PATIENT DATA<br />

Function Range<br />

Shows the calculated value <strong>of</strong><br />

ventilator breathing circuit<br />

pressure over an entire<br />

respiratory cycle. Updated at the<br />

beginning <strong>of</strong> each breath.<br />

Shows the maximum pressure<br />

measured during inspiration.<br />

Updated at the beginning <strong>of</strong><br />

each expiratory phase. (Default<br />

pressure display.)<br />

Shows the pressure measured at<br />

the end <strong>of</strong> the plateau period <strong>of</strong><br />

a mandatory inspiration<br />

(whether the inspiration is in a<br />

regular VCV breath or is part <strong>of</strong><br />

an inspiratory pause maneuver).<br />

Updated at the beginning <strong>of</strong><br />

each expiratory phase. The<br />

PRESSURE display shows a blank<br />

if the ventilator does not detect<br />

a stable plateau pressure.<br />

Shows the pressure measured at<br />

the expiratory limb before any<br />

inspiratory effort. Updated at<br />

the beginning <strong>of</strong> each<br />

inspiratory phase.<br />

0 to 99 cmH 2O<br />

(0 to 99 hPa)<br />

Accuracy:<br />

± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />

0 to 140 cmH 2O<br />

(0 to 140 hPa)<br />

Accuracy:<br />

± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />

0 to 140 cmH 2O<br />

(0 to 140 hPa)<br />

Accuracy:<br />

± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />

0 to 140 cmH 2O<br />

(0 to 140 hPa)<br />

Accuracy:<br />

± (3 + 4% <strong>of</strong> reading) cmH 2 O<br />

1�24 10066984 Rev. D 700 Series Ventilator Operator’s Manual