Setting up the ventilator<br />

NOTE:<br />

• <strong>Covidien</strong> recommends that you run SST every 15 days,<br />

between patients, and when you change the ventilator<br />

breathing circuit. <strong>Covidien</strong> recognizes that the protocol for<br />

running SST varies widely among healthcare institutions.<br />

It is not possible for <strong>Covidien</strong> to specify or require specific<br />

practices that will meet all needs, or to be responsible for<br />

the effectiveness <strong>of</strong> those practices.<br />

• The ventilator uses an oxygen sensor to trigger an alarm if<br />

the delivered % O2 is 10 percentage points above or below<br />

the % O2 setting. You can view the ventilator’s % O2 measurement by using the User settings function <strong>of</strong> the<br />

MENU key.<br />

• Check the inspiratory and expiratory limbs <strong>of</strong> the ventilator<br />

breathing circuit and the collector vial and in-line water<br />

traps regularly for water buildup. Under certain conditions,<br />

they can fill quickly. In case <strong>of</strong> a partial occlusion alarm,<br />

check for water in the ventilator breathing circuit. In very<br />

severe cases, such a water occlusion can potentially cause<br />

the ventilator to enter occlusion cycling mode. Empty and<br />

clean the collector vial and in-line water traps as necessary.<br />

• <strong>Covidien</strong> recommends using pediatric circuits when<br />

ventilating patients with 5-mm or smaller internal-diameter<br />

artificial airways.<br />

Figure 2-8 shows how to connect the ventilator breathing circuit,<br />

including the inspiratory filter, humidifier (if used), inspiratory limb,<br />

patient wye, expiratory limb, collector vial, and expiratory filter.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 2�11