Introduction<br />
<strong>Table</strong> 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS<br />
Key/indicator Specifies... Range<br />
Mandatory (PCV) settings (760 Ventilator only) (continued)<br />
INSPIRATORY<br />
PRESSURE<br />
(760 only)<br />
T I /I:E RATIO<br />
(760 only)<br />
RISE TIME<br />
FACTOR<br />
(760 only)<br />
Pressure above PEEP during<br />
the inspiratory phase <strong>of</strong> a PCV<br />
breath.<br />
You can use the MENU key to<br />
select inspiratory time (T I ) or<br />
I:E ratio as the breath timing<br />
setting for a PCV breath. You<br />
can change the selected<br />
breath timing setting (T I or I:E<br />
ratio), but the setting remains<br />
constant when you change<br />
the respiratory rate in PCV.<br />
Selecting I:E ratio makes the<br />
set ratio <strong>of</strong> inspiratory time to<br />
expiratory time for a PCV<br />
breath.<br />
The time for inspiratory<br />
pressure to rise from 0 to 95%<br />
<strong>of</strong> the target pressure level<br />
during a PCV breath.<br />
A setting <strong>of</strong> 100 = a 100-ms<br />
rise time, and a setting <strong>of</strong><br />
5 = 80% <strong>of</strong> the inspiratory<br />
time or 2500 ms, whichever is<br />
less.<br />
When you adjust this setting,<br />
the message window shows<br />
the actual time (in seconds) to<br />
reach 95% <strong>of</strong> target pressure.<br />
5 to 80 cmH 2 O (5 to 80 hPa)<br />
Accuracy:<br />
± (3 + 2.5% <strong>of</strong> setting) cmH 2O<br />
Inspiratory time (T I ):<br />
0.2 to 8 seconds<br />
Accuracy: ± 0.05 second<br />
I:E ratio: ≤ 4:1<br />
Accuracy: ± (0.1 + 2%)<br />
5 to 100<br />
NOTE:<br />
To help determine the correct setting for inspiratory time, during<br />
PCV the message window displays peak inspiratory flow, end<br />
inspiratory flow, and end exhalation flow in L/min.<br />
700 Series Ventilator Operator’s Manual 10066984 Rev. D 1�17