31.10.2012 Views

Table of Contents - Covidien

Table of Contents - Covidien

Table of Contents - Covidien

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

Self tests (SST and EST)<br />

SECTION<br />

All 700 Series Ventilators include these built-in self tests:<br />

• Short self test (SST) is a short (about 2 to 3 minutes) and simple<br />

sequence <strong>of</strong> tests that verifies proper ventilator operation, checks<br />

the ventilator breathing circuit for leaks, and measures the circuit<br />

compliance. Power on self test (POST) is part <strong>of</strong> SST.<br />

• Extended self test (EST) is a thorough test <strong>of</strong> the operational<br />

integrity <strong>of</strong> the ventilator, both the electronics and the<br />

pneumatics. POST and SST are part <strong>of</strong> EST.<br />

Warning<br />

• Before running SST or EST, you must disconnect the<br />

ventilator from the patient. Running SST or EST while the<br />

ventilator is connected to the patient can injure the patient.<br />

• A fault identified in SST or EST indicates that the ventilator or<br />

an associated component is defective. A defective ventilator<br />

or associated component should be repaired before the<br />

ventilator is returned to service, unless it can be determined<br />

with certainty that the defect cannot create a hazard for the<br />

patient, or add to the risks which may arise from other<br />

hazards.