Test number<br />

and name<br />

9. I/E filter test Checks pressure drop across<br />

inspiratory and expiratory<br />

limbs <strong>of</strong> entire patient<br />

system. Checks pressure<br />

drop across filters.<br />

Disconnect and reconnect<br />

inspiration (I) tube from/to<br />

inspiratory filter outlet, and<br />

exhalation (E) tube from/to<br />

expiratory filter inlet, as<br />

directed. Block and unblock<br />

wye with a no. 2 stopper, as<br />

directed. Press ACCEPT to<br />

signify you have performed<br />

each requested action.<br />

10. PEEP<br />

system test<br />

11. Check<br />

valve test<br />

<strong>Table</strong> 4-6: EST test sequence (continued)<br />

Verifies that PEEP system can<br />

generate and maintain<br />

preset PEEP levels within<br />

either <strong>of</strong> two sets <strong>of</strong> limits. If<br />

results are within the inner<br />

set <strong>of</strong> limits, test passes the<br />

first time. If the results are<br />

outside <strong>of</strong> the outer set <strong>of</strong><br />

limits, test fails the first<br />

time. If the results are<br />

outside the inner limits but<br />

within the outer limits, the<br />

ventilator adjusts its<br />

calibration table in<br />

nonvolatile RAM (NVRAM)<br />

and repeats the test to verify<br />

the calibration. The test fails<br />

if the calibration cannot be<br />

verified after five attempts.<br />

Self tests (SST and EST)<br />

Function Comments<br />

Checks check valves on inlet<br />

and outlet ports <strong>of</strong> cylinder<br />

and on exhalation limb <strong>of</strong><br />

ventilator breathing circuit.<br />

Fault if not passed. Also in<br />

SST. Mandatory in EST, but<br />

optional in SST.<br />

NOTE:<br />

Overriding this fault<br />

could result in<br />

inadequate bacteria<br />

protection or excessive<br />

resistance to inspiration<br />

or exhalation.<br />

Failure if not passed. Also in<br />

SST, but not identical.<br />

Failure if not passed. Not in<br />

SST.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 4-19