Specifications<br />
C.1 Physical<br />
<strong>Table</strong> C-1: Physical specifications<br />
Weight Ventilator only: 30 kg (66 lb)<br />
Cart only: 18 kg (40 lb)<br />
External battery: 12.75 kg (28 lb)<br />
Dimensions Ventilator only: 378 mm H x 515 mm W x<br />
370 mm D<br />
(14.9 in. H x 20.3 in. W x<br />
14.6 in. D)<br />
Ventilator and cart: 1255 mm H x 515 mm W x<br />
370 mm D<br />
(49.4 in. H x 20.3 in. W x<br />
14.6 in. D)<br />
Connectors Inspiratory limb connector: ISO 22-mm conical male<br />
Expiratory limb connector: ISO 22-mm conical female<br />
Gas exhaust port: ISO 30-mm conical<br />
Oxygen inlet: DISS, DISS female, NIST, Air<br />
Liquide*, or Australian<br />
type (depending on<br />
country and configuration)<br />
Ventilator<br />
heat<br />
generation<br />
400 to 500 British thermal units (BTUs) per hour<br />
Oxygen<br />
sensor life<br />
Gas mixing<br />
system<br />
Two years, nominal. (Actual life depends on operating<br />
environment; operation at higher temperature or FIO2 levels will result in shorter sensor life.)<br />
Range <strong>of</strong> flow delivery to the patient: Up to 300 L/min<br />
for pressure support ventilation (PSV) spontaneous<br />
breaths or pressure control ventilation (PCV)<br />
mandatory breaths; 3 to 150 L/min for mandatory<br />
or assisted volume control ventilation (VCV)<br />
breaths.<br />
Leakage from one gas system to another:<br />
Not applicable<br />
(no high-pressure air source).<br />
Design pressure: 50 psi (345 kPa)<br />
Operating pressure range: 40 to 90 psi (275 to 620 kPa)<br />
Oxygen regulator bleed: Up to 3 L/min<br />
C-2 10066984 Rev. D 700 Series Ventilator Operator’s Manual