Self tests (SST and EST)<br />

Test number<br />

and name<br />

7. Leak test Checks ventilator breathing<br />

circuit for leaks.<br />

7. P e P cyl<br />

compare<br />

(also part<br />

<strong>of</strong> test 7)<br />

7. Auto-zero<br />

sol test (also<br />

part <strong>of</strong><br />

test 7)<br />

8. Circuit comp<br />

test<br />

<strong>Table</strong> 4-6: EST test sequence (continued)<br />

Function Comments<br />

Verifies proper functioning<br />

<strong>of</strong> pressure transducers and<br />

ventilator breathing circuit.<br />

Verifies the proper function<br />

<strong>of</strong> the P e and P cyl solenoids.<br />

Determines ventilator<br />

breathing circuit<br />

compliance.<br />

Fault if not passed. Also in<br />

SST.<br />

NOTE:<br />

Overriding this fault<br />

could cause improper<br />

compliance<br />

compensation,<br />

inaccurate tidal volume<br />

delivery, or autocycling.<br />

Fault if not passed. Also in<br />

SST.<br />

NOTE:<br />

Overriding this fault<br />

could cause improper<br />

compliance<br />

compensation,<br />

inaccurate tidal volume<br />

delivery, or autocycling.<br />

Failure if not passed. Also in<br />

SST.<br />

NOTE:<br />

Overriding this fault<br />

could cause improper<br />

compliance<br />

compensation,<br />

inaccurate tidal volume<br />

delivery, or autocycling.<br />

Fault if not passed. Also in<br />

SST.<br />

4-18 10066984 Rev. D 700 Series Ventilator Operator’s Manual