Introduction<br />

<strong>Table</strong> 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS<br />

Key/indicator Specifies... Range<br />

Other keys, knobs, and indicators (continued)<br />

EXP PAUSE<br />

(760 only)<br />

Allows you to measure the patient’s auto-PEEP. An EXP PAUSE<br />

maneuver causes the ventilator to close the exhalation valve<br />

at the end <strong>of</strong> the expiratory phase, and not deliver the next<br />

mandatory breath. At the end <strong>of</strong> the maneuver, the message<br />

window shows the calculated value for auto-PEEP (expiratory<br />

pressure at the beginning <strong>of</strong> the maneuver minus expiratory<br />

pressure at the end <strong>of</strong> the maneuver) and total PEEP for<br />

30 seconds.<br />

The message window shows the end expiratory flow in L/min<br />

at the beginning <strong>of</strong> each breath. If there is expiratory flow<br />

when the ventilator delivers the next breath, auto-PEEP is<br />

present.<br />

The EXP PAUSE continues as long as you hold down the key,<br />

and should last only until expiratory pressure stabilizes. An<br />

EXP PAUSE maneuver ends when you release the key, the<br />

patient initiates a breath, an alarm occurs, the expiratory<br />

phase (including the maneuver) lasts more than 20 seconds,<br />

or the ventilator detects a leak.<br />

Auto-PEEP:<br />

Range: 1 to 35 cmH 2 O.<br />

Accuracy: ± (1 cmH 2O + 3% <strong>of</strong> reading).<br />

The EXP PAUSE maneuver is unavailable when the<br />

RESPIRATORY RATE setting is less than 3 /minute.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 1�21