When you<br />
see this<br />
message...<br />
DELIV GAS<br />
LOW TEMP<br />
EXH CCT<br />
HI TEMP<br />
EXH CCT<br />
LO TEMP<br />
FAN FAILED<br />
ALERT<br />
<strong>Table</strong> 7-2: Technical alarms (continued)<br />
It means... Do this...<br />
High-priority alarm.<br />
Inspiration temperature too low.<br />
Autoreset when temperature is<br />
within acceptable limit.<br />
High-priority alarm.<br />
Temperature <strong>of</strong> the exhalation<br />
heater is above acceptable limit.<br />
The ventilator continues to<br />
annunciate the alarm while the<br />
condition persists.<br />
Autoreset when temperature is<br />
within acceptable limit.<br />
High-priority alarm.<br />
Temperature <strong>of</strong> the exhalation<br />
heater is below acceptable limit.<br />
The ventilator continues to<br />
annunciate the alarm while the<br />
condition persists.<br />
Autoreset when temperature is<br />
within acceptable limit.<br />
High-priority alarm.<br />
Fan filter occluded or fan not<br />
operational.<br />
Alarm does not autoreset; you<br />
must press the alarm reset key.<br />
Alarm handling<br />
Make sure room<br />
temperature is not too<br />
low.<br />
Remove ventilator from<br />
use and contact service.<br />
If problem persists,<br />
remove ventilator from<br />
use and contact service.<br />
Be aware that spirometry<br />
may be affected<br />
(monitored values may<br />
be higher than actual).<br />
If problem persists,<br />
remove ventilator from<br />
use and contact service.<br />
Be aware that spirometry<br />
may be affected<br />
(monitored values may<br />
be lower than actual).<br />
Make sure ventilator has<br />
warmed up sufficiently.<br />
Check fan filter for<br />
occlusions and clean if<br />
necessary. (Appendix A<br />
tells you how to check<br />
and clean the fan filter.)<br />
Press the alarm reset key.<br />
If this does not resolve<br />
the problem, contact<br />
service.<br />
700 Series Ventilator Operator’s Manual 10066984 Rev. D 7-15