When you<br />

see this<br />

message...<br />

DELIV GAS<br />

LOW TEMP<br />

EXH CCT<br />

HI TEMP<br />

EXH CCT<br />

LO TEMP<br />

FAN FAILED<br />

ALERT<br />

<strong>Table</strong> 7-2: Technical alarms (continued)<br />

It means... Do this...<br />

High-priority alarm.<br />

Inspiration temperature too low.<br />

Autoreset when temperature is<br />

within acceptable limit.<br />

High-priority alarm.<br />

Temperature <strong>of</strong> the exhalation<br />

heater is above acceptable limit.<br />

The ventilator continues to<br />

annunciate the alarm while the<br />

condition persists.<br />

Autoreset when temperature is<br />

within acceptable limit.<br />

High-priority alarm.<br />

Temperature <strong>of</strong> the exhalation<br />

heater is below acceptable limit.<br />

The ventilator continues to<br />

annunciate the alarm while the<br />

condition persists.<br />

Autoreset when temperature is<br />

within acceptable limit.<br />

High-priority alarm.<br />

Fan filter occluded or fan not<br />

operational.<br />

Alarm does not autoreset; you<br />

must press the alarm reset key.<br />

Alarm handling<br />

Make sure room<br />

temperature is not too<br />

low.<br />

Remove ventilator from<br />

use and contact service.<br />

If problem persists,<br />

remove ventilator from<br />

use and contact service.<br />

Be aware that spirometry<br />

may be affected<br />

(monitored values may<br />

be higher than actual).<br />

If problem persists,<br />

remove ventilator from<br />

use and contact service.<br />

Be aware that spirometry<br />

may be affected<br />

(monitored values may<br />

be lower than actual).<br />

Make sure ventilator has<br />

warmed up sufficiently.<br />

Check fan filter for<br />

occlusions and clean if<br />

necessary. (Appendix A<br />

tells you how to check<br />

and clean the fan filter.)<br />

Press the alarm reset key.<br />

If this does not resolve<br />

the problem, contact<br />

service.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 7-15