C.5 Technical<br />

Maximum<br />

limited<br />

pressure<br />

Maximum<br />

ventilating<br />

pressure<br />

Measuring<br />

and display<br />

devices<br />

<strong>Table</strong> C-5: Technical specifications<br />

Specifications<br />

92 cmH 2 O (92 hPa).<br />

A dedicated backup circuit opens the safety valve if<br />

system pressure exceeds 115 cmH 2 O (113 hPa).<br />

89 cmH 2 O (89 hPa) for VCV breaths or 80 cmH 2 O<br />

(80 hPa) for PSV breaths, ensured by HIGH PRESSURE<br />

limit.<br />

Purpose: Mean airway pressure<br />

Sensing position: Exhalation limb<br />

Range: 0 to 99 cmH2O (0 to 99 hPa)<br />

Type: Silicon solid state<br />

Purpose: Peak pressure<br />

Sensing position: Exhalation limb<br />

Range: 0 to 140 cmH2O (0 to 140 hPa)<br />

Type: Silicon solid state<br />

Purpose: Plateau pressure (760 Ventilator<br />

only)<br />

Sensing position: Exhalation limb<br />

Range: 0 to 140 cmH2O (0 to 99 hPa)<br />

Type: Silicon solid state<br />

Purpose: PEEP/CPAP pressure<br />

Sensing position: Exhalation limb<br />

Range: 0 to 140 cmH2O (0 to 140 hPa)<br />

Type: Silicon solid state<br />

Purpose: Rate<br />

Sensing position: Microprocessor<br />

Range: 3 to 500/min<br />

Type: Calculated from inspiratory and<br />

expiratory time measurements<br />

Purpose: I:E ratio<br />

Sensing position: Microprocessor<br />

Range: 1:99.9 to 1:9.9<br />

Type: Calculated from inspiratory and<br />

expiratory time measurements<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D C-7