The MENU key<br />

6.4 User settings<br />

1. Once you’ve selected the User settings menu function, press<br />

ACCEPT. Turn the knob to display one <strong>of</strong> these messages:<br />

Endotracheal tube<br />

Humidifier type<br />

Date and time set<br />

Apnea interval (Ta)<br />

VCV flow pattern<br />

Speaking valve setup<br />

Alarm volume<br />

PCV timing setting<br />

Volume LED bar<br />

2. Press ACCEPT to select the option.<br />

6.4.1 Endotracheal tube<br />

If you select Endotracheal tube, the ventilator displays the current ET<br />

size setting.<br />

1. Turn the knob to select the appropriate size for the patient.<br />

2. Press ACCEPT to apply and return to the list <strong>of</strong> user settings.<br />

Press CLEAR to exit without making any changes.<br />

6.4.2 Humidifier type<br />

1. If you select Humidifier type, the ventilator displays the<br />

currently-selected type:<br />

• HME: Heat-moisture exchanger (“artificial nose”)<br />

• Dual heated wire: humidifier with heated wire on expiratory<br />

limb, or on both inspiratory and expiratory limbs<br />

• No heated wire: conventional humidifier, no heated wire on<br />

expiratory limb<br />

2. Turn the knob to select the humidifier type attached, then press<br />

ACCEPT. Press CLEAR to exit without making any changes.<br />

6-8 10066984 Rev. D 700 Series Ventilator Operator’s