Self tests (SST and EST)<br />

Test number<br />

and name<br />

12. Piston leak<br />

test<br />

13. Lightsdisplays<br />

test<br />

Checks piston leak against a<br />

table <strong>of</strong> leak values stored<br />

in NVRAM.<br />

First remove inspiratory filter;<br />

then block and unblock<br />

inspiratory filter port as<br />

directed, using no. 3 stopper.<br />

Press ACCEPT to signify you<br />

have performed each<br />

requested action. Replace<br />

inspiratory tubing at end <strong>of</strong><br />

test.<br />

Verifies that lights and<br />

displays function.<br />

Observe each named display,<br />

and press ACCEPT to verify<br />

that it is lit. Press CLEAR if a<br />

display is not lit.<br />

14. Keys test Verifies that every row and<br />

column <strong>of</strong> key matrix can be<br />

correctly read.<br />

Press each named key (each<br />

key’s LED will also be lit).<br />

Press CLEAR if a key’s LED is<br />

not lit.<br />

15. Main alarm<br />

test<br />

<strong>Table</strong> 4-6: EST test sequence (continued)<br />

Function Comments<br />

Checks whether various<br />

alarm sounds are audible (at<br />

maximum volume).<br />

Listen for each named<br />

sound, then press ACCEPT to<br />

verify that it is audible. Press<br />

CLEAR if a sound is not<br />

audible.<br />

Failure if not passed. Not in<br />

SST.<br />

Failure if not passed. Not in<br />

SST.<br />

Failure if not passed. Not in<br />

SST.<br />

Failure if not passed. Also in<br />

SST.<br />

4-20 10066984 Rev. D 700 Series Ventilator Operator’s Manual