Table of Contents - Covidien
Table of Contents - Covidien
Table of Contents - Covidien
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Self tests (SST and EST)<br />
Test number<br />
and name<br />
12. Piston leak<br />
test<br />
13. Lightsdisplays<br />
test<br />
Checks piston leak against a<br />
table <strong>of</strong> leak values stored<br />
in NVRAM.<br />
First remove inspiratory filter;<br />
then block and unblock<br />
inspiratory filter port as<br />
directed, using no. 3 stopper.<br />
Press ACCEPT to signify you<br />
have performed each<br />
requested action. Replace<br />
inspiratory tubing at end <strong>of</strong><br />
test.<br />
Verifies that lights and<br />
displays function.<br />
Observe each named display,<br />
and press ACCEPT to verify<br />
that it is lit. Press CLEAR if a<br />
display is not lit.<br />
14. Keys test Verifies that every row and<br />
column <strong>of</strong> key matrix can be<br />
correctly read.<br />
Press each named key (each<br />
key’s LED will also be lit).<br />
Press CLEAR if a key’s LED is<br />
not lit.<br />
15. Main alarm<br />
test<br />
<strong>Table</strong> 4-6: EST test sequence (continued)<br />
Function Comments<br />
Checks whether various<br />
alarm sounds are audible (at<br />
maximum volume).<br />
Listen for each named<br />
sound, then press ACCEPT to<br />
verify that it is audible. Press<br />
CLEAR if a sound is not<br />
audible.<br />
Failure if not passed. Not in<br />
SST.<br />
Failure if not passed. Not in<br />
SST.<br />
Failure if not passed. Not in<br />
SST.<br />
Failure if not passed. Also in<br />
SST.<br />
4-20 10066984 Rev. D 700 Series Ventilator Operator’s Manual