Introduction<br />

<strong>Table</strong> 1-3: 700 Series Ventilator keyboards: VENTILATOR SETTINGS<br />

Key/indicator Specifies... Range<br />

Spontaneous (PSV) settings<br />

SUPPORT<br />

PRESSURE<br />

RISE TIME<br />

FACTOR<br />

(760 only)<br />

Pressure above PEEP<br />

maintained during<br />

spontaneous inspiration.<br />

Support pressure is<br />

terminated when inspiratory<br />

flow falls to 25% <strong>of</strong> peak<br />

inspiratory flow, or to the<br />

exhalation sensitivity setting<br />

(760 only), or 10 L/min or 25%<br />

<strong>of</strong> peak flow, whichever is<br />

lower (740 only). Maximum<br />

inspiratory time is 3.5 seconds<br />

for adults, and 2.5 seconds for<br />

pediatric patients.<br />

The time for inspiratory<br />

pressure to rise from 0 to 95%<br />

<strong>of</strong> the target pressure level<br />

during a PSV breath.<br />

A setting <strong>of</strong> 100 = a 100-ms<br />

rise time, and a setting <strong>of</strong><br />

5 = 80% <strong>of</strong> the inspiratory<br />

time or 1500 ms (when adult<br />

ventilator breathing circuit is<br />

selected) or 600 ms (when<br />

pediatric circuit is selected),<br />

whichever is less.<br />

When you adjust this setting,<br />

the message window shows<br />

the actual time (in seconds) to<br />

reach 95% <strong>of</strong> target pressure.<br />

0 to 70 cmH 2 O (0 to 70 hPa)<br />

Accuracy:<br />

± (3 + 2.5% <strong>of</strong> setting) cmH 2O<br />

5 to 100<br />

1�18 10066984 Rev. D 700 Series Ventilator Operator’s Manual