Alarm handling<br />

When you<br />

see this<br />

message...<br />

LOW EXT<br />

BATTERY<br />

LOW INT<br />

BATTERY<br />

LOW O 2<br />

SUPPLY<br />

<strong>Table</strong> 7-2: Technical alarms (continued)<br />

It means... Do this...<br />

High-priority alarm.<br />

Low external battery voltage<br />

detected during POST, although<br />

ac power is sufficient. This alarm<br />

occurs only when the internal<br />

battery is also low or missing.<br />

Autoreset after 2 minutes, or<br />

when external battery is<br />

replaced. Once reset, alarm is not<br />

reannunciated.<br />

High-priority alarm.<br />

Low internal battery voltage<br />

detected during POST, although<br />

ac power is sufficient.<br />

Autoreset after 2 minutes. Once<br />

reset, alarm is not<br />

reannunciated.<br />

High-priority alarm.<br />

Low oxygen supply pressure.<br />

Ventilation continues with<br />

reduced % O 2 . (Alarm is not<br />

activated when % O 2 setting is<br />

21%.)<br />

Autoreset when sufficient<br />

oxygen supply pressure is<br />

detected.<br />

Verify reliable ac power<br />

source.<br />

Check battery<br />

connections.<br />

Replace external battery<br />

if necessary.<br />

Verify reliable ac<br />

power source.<br />

Contact service to check<br />

internal battery<br />

connections and replace<br />

internal battery if<br />

necessary.<br />

Check patient.<br />

Check integrity <strong>of</strong><br />

oxygen supply and<br />

connections. Increase<br />

oxygen supply pressure if<br />

necessary.<br />

7-18 10066984 Rev. D 700 Series Ventilator Operator’s Manual