Results <strong>of</strong><br />

ventilator<br />

breathing<br />

system<br />

testing<br />

(using<br />

circuits<br />

identified<br />

for use with<br />

700 Series<br />

Ventilators)<br />

<strong>Table</strong> C-5: Technical specifications (continued)<br />

Specifications<br />

NOTE:<br />

To achieve compliance with the more rigid<br />

requirements for oxygen monitoring (as set forth<br />

in ISO 7767), <strong>Covidien</strong> recommends using an<br />

external monitor that meets that standard.<br />

Internal volume: Not applicable; 700 Series Ventilators<br />

automatically adjust for volume losses due to gas<br />

compressibility (that is, automatic compliance<br />

compensation), subject to a maximum delivered<br />

volume <strong>of</strong> 2.5 L.<br />

NOTE:<br />

To ensure that compliance compensation<br />

functions correctly, the user must run SST with<br />

the circuit configured as intended for use on<br />

the patient.<br />

NOTE:<br />

Ventilator breathing circuit testing specifications<br />

are based on the recommended configurations<br />

shown in Figure C-1 (heated wire humidifier<br />

without water traps and non-heated wire<br />

humidifier with water traps).<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D C-9