C.2 Environmental<br />

Specifications<br />

Inspiratory and expiratory filters (disposable and<br />

reusable): 99.97% retention <strong>of</strong> 0.3 μm particles at<br />

100 L/min.<br />

Minimum (level 1): 74 dBA at 1 m<br />

Maximum (level 5): 88 dBA at 1 m<br />

<strong>Table</strong> C-2: Environmental specifications<br />

Temperature Operating: 5 to 45 °C (41 to 113 °F)<br />

at 10 to 95% relative humidity<br />

Storage: -40 to 60 °C (-40 to 140 °F) at<br />

10 to 95% relative humidity<br />

Atmospheric<br />

pressure<br />

<strong>Table</strong> C-1: Physical specifications (continued)<br />

Operating: 8.7 to 16.0 psi<br />

(600 to 1100 hPa)<br />

Storage: 7.3 to 16.0 psi<br />

(500 to 1100 hPa)<br />

Altitude Operating: Up to 4570 m (15,000 ft)<br />

Storage: Up to 15,240 m (50,000 ft)<br />

Oxygen inlet<br />

supply<br />

Pressure:<br />

Warning<br />

40 to 90 psi (275 to 620 kPa)<br />

Due to excessive restriction <strong>of</strong> certain hose<br />

assemblies (listed in <strong>Table</strong> B-1), reduced FIO 2<br />

levels may result when oxygen inlet pressures<br />

< 50 psi (345 kPa) are employed. Make sure<br />

oxygen inlet pressure is ≥ 50 psi (345 kPa) when<br />

using these hose assemblies, to maintain correct<br />

FIO 2 levels.<br />

Flow: At least 160 L/min at standard<br />

temperature and pressure,<br />

dry (STPD) required<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D C-3