The MENU key<br />

8. If you choose to disable the speaking valve option, the ventilator<br />

displays the following messages to tell you that it is resuming<br />

exhaled volume monitoring to detect the LOW TIDAL VOLUME<br />

and HIGH TIDAL VOLUME alarms, and restoring the previous<br />

settings for PEEP/CPAP, LOW MINUTE VOLUME alarm, DISCONNECT<br />

alarm, LOW INSP PRESSURE alarm, and the apnea interval:<br />

Volume alarms based<br />

on exhaled volume<br />

Restore to x.xL<br />

Low minute volume<br />

Alarm RESTORED<br />

Disconnect alarm<br />

Restore to x.xcmH2O PEEP/CPAP<br />

Restore to xx s<br />

Apnea interval<br />

When A/C or SIMV mode is active:<br />

Restore to xx cmH2O Low insp pressure<br />

When SPONT mode is active:<br />

No LIP alrm in SPONT<br />

NOTE:<br />

If a setting changed when the speaking valve option was<br />

enabled that makes any <strong>of</strong> these previous settings invalid, the<br />

ventilator displays the maximum valid value for that setting in<br />

its Restore message.<br />

On the 760 Ventilator, if the volume bar graph was disabled by<br />

turning the speaking valve option on, it is re-enabled when<br />

the option is <strong>of</strong>f.<br />

Press ACCEPT at each prompt to disable the option and return<br />

to user settings. Press CLEAR to exit and leave the option<br />

enabled.<br />

6-16 10066984 Rev. D 700 Series Ventilator Operator’s