Maintenance<br />

Replace reusable filters after a maximum <strong>of</strong> one year <strong>of</strong> service or<br />

100 autoclave cycles, whichever comes first. When you put a reusable<br />

filter into service, write the anticipated replacement date on the<br />

filter and keep <strong>of</strong> record <strong>of</strong> the number <strong>of</strong> autoclave cycles to which<br />

the filter has been subjected. Discard disposable filters between<br />

patients.<br />

A.2.2 Daily or as required: collector vial<br />

At every ventilator breathing circuit change, autoclave or disinfect<br />

the collector vial. The patented design <strong>of</strong> the collector vial maintains<br />

circuit pressure (minimizes PEEP loss) and prevents condensate from<br />

spraying when you remove the collector vial from the ventilator<br />

breathing circuit.<br />

Empty the collector vial before it fills:<br />

1. Turn the collector vial counterclockwise to release it from the<br />

ventilator breathing circuit.<br />

2. Empty the collector vial, referring to Figure A-1.<br />

3. Replace the empty collector vial: turn clockwise to secure the<br />

vial into place, which reopens its access to the ventilator<br />

breathing circuit.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D A-9