Self tests (SST and EST)<br />

Safety valve<br />

test<br />

PEEP system<br />

test<br />

Flow/O 2<br />

sensor test<br />

<strong>Table</strong> 4-2: SST sequence <strong>of</strong> tests (continued)<br />

Test Function Comments<br />

Verifies that safety valve<br />

relieves excess circuit<br />

pressure. Displays<br />

expiratory pressure on<br />

pressure bar graph.<br />

Verifies that the PEEP<br />

system can generate and<br />

maintain preset PEEP levels.<br />

Displays expiratory pressure<br />

on pressure bar graph.<br />

Checks the accuracy <strong>of</strong> the<br />

exhalation flow sensor and<br />

the oxygen sensor.<br />

I/E filter test Checks the pressure drop<br />

across the inspiratory and<br />

expiratory limbs <strong>of</strong> the<br />

entire patient system.<br />

Checks the pressure drop<br />

across the filters.<br />

Failure if not passed.<br />

Failure if not passed.<br />

Failure if flow sensor test<br />

does not pass.<br />

Fault if oxygen sensor test<br />

does not pass. Use the<br />

MENU key’s Oxygen sensor<br />

function to recalibrate the<br />

oxygen sensor (see<br />

Chapter 6).<br />

Overriding a fault (oxygen<br />

sensor test only) could result<br />

in inaccurate % O 2<br />

measurement.<br />

Fault if not passed. This test<br />

is optional: pressing CLEAR<br />

at the initial prompt skips<br />

this test.<br />

Skipping this test or<br />

overriding a fault could<br />

result in inadequate<br />

bacteria protection or<br />

excessive resistance to<br />

inspiration or exhalation.<br />

4-10 10066984 Rev. D 700 Series Ventilator Operator’s Manual