Alarm handling<br />

The highest-priority active alarm always flashes on the first line <strong>of</strong><br />

the message window. If no alarm is active, the first line displays the<br />

highest-priority autoreset alarm. You can use the menu function to<br />

view all remaining active and autoreset alarms. (Chapter 6 <strong>of</strong> this<br />

manual tells you how to use the menu function.) Figure 7-1<br />

summarizes how to view active alarms.<br />

7-00054<br />

ALARM SETTINGS<br />

7-00075<br />

7.1 Autoreset alarms<br />

- +<br />

ALARM<br />

Red ALARM indicator flashes when<br />

high-priority alarm is active.<br />

CAUTION<br />

Yellow CAUTION indicator flashes<br />

when medium-priority alarm is active.<br />

Figure 7-1. Viewing active alarms<br />

Silences alarm sounds but<br />

does not change visual<br />

indicators.<br />

The highest-priority active<br />

alarm flashes on the first line<br />

<strong>of</strong> the message window.<br />

If the active alarm is an operatoradjustable<br />

alarm, its key light flashes and<br />

the ALARM SETTINGS window is blank.<br />

When an alarm is autoreset (that is, reset because the condition<br />

corrected itself), it is added to the Autoreset alarms list. You can view<br />

the Autoreset alarms list using the MENU key. (Chapter 6 tells you<br />

how to use the MENU key.)<br />

When an alarm autoresets, the CAUTION or ALARM indicator lights<br />

steadily (instead <strong>of</strong> flashing). Any alarm key lights that were flashing<br />

when the alarm was active return to their previous states (<strong>of</strong>f or<br />

steadily lit).<br />

Pressing alarm reset immediately clears all autoreset alarms. For accurate<br />

alarm condition evaluation, press alarm silence and view autoreset<br />

alarms (using the Autoreset alarms function, as described in Chapter 6).<br />

7-2 10066984 Rev. D 700 Series Ventilator Operator’s Manual