The MENU key<br />

6.4.4 Apnea interval (Ta)<br />

1. If you select Apnea inteval (Ta), the ventilator displays this<br />

message:<br />

xx (10 - 60) s<br />

Apnea interval<br />

where xx is the apnea interval in seconds.<br />

2. Turn the knob to select an apnea interval from 10 to 60 seconds.<br />

3. Press CLEAR to cancel any changes, or press ACCEPT to apply the<br />

new interval and return to the list <strong>of</strong> user settings.<br />

6.4.5 VCV flow pattern<br />

1. If you select VCV flow pattern, the ventilator displays this<br />

message:<br />

VCV Flow: xxxxxx<br />

Ti = y.yy s<br />

where xxxxxx is the currently selected flow pattern and y.yy is<br />

the current inspiratory time (the value <strong>of</strong> y.yy depends on the<br />

selected flow pattern).<br />

2. Turn the knob to select RAMP (descending ramp flow pattern)<br />

or SQUARE (square flow pattern).<br />

If the flow pattern change would result in a violation <strong>of</strong> breath<br />

delivery boundaries, the ventilator displays one <strong>of</strong> these<br />

messages:<br />

Change not permitted<br />

Adjust VCV first<br />

or<br />

Change not permitted<br />

Adjust Ap. VCV first<br />

3. Press CLEAR to cancel any changes, or press ACCEPT to select the<br />

new flow pattern and return to the list <strong>of</strong> user settings.<br />

6-10 10066984 Rev. D 700 Series Ventilator Operator’s