Volume (continued)<br />
SPONT<br />
MINUTE<br />
VOLUME<br />
(L)<br />
(760 only)<br />
Shows the patient’s measured<br />
expiratory minute volume for all<br />
spontaneous breaths, based on<br />
the previous 60 seconds or 8<br />
breaths (whichever interval is<br />
shorter). Updated at the<br />
beginning <strong>of</strong> each breath.<br />
The calculation is reset when<br />
ventilation starts, when apnea<br />
ventilation starts or autoresets,<br />
when you change the mode or<br />
breath type, and when you press<br />
the alarm reset key.<br />
Alarm settings<br />
HIGH RATE An active alarm indicates that<br />
measured respiratory rate is<br />
higher than the alarm setting.<br />
HIGH<br />
TIDAL<br />
VOLUME<br />
LOW INSP<br />
PRESSURE<br />
LOW<br />
TIDAL<br />
VOLUME<br />
<strong>Table</strong> 1-4: 700 Series Ventilator keyboards: PATIENT DATA<br />
Key/<br />
indicator<br />
Function Range<br />
An active alarm indicates that<br />
exhaled volume for three out <strong>of</strong><br />
four consecutive breaths was<br />
above the alarm setting.<br />
An active alarm indicates that<br />
monitored circuit pressure is<br />
below the alarm setting at the<br />
end <strong>of</strong> inspiration. Inactive in for<br />
any spontaneous breath.<br />
An active alarm indicates that<br />
delivered volume for three out<br />
<strong>of</strong> four consecutive breaths were<br />
below the alarm setting. (If this<br />
alarm is set to 0 ml and breath<br />
type is PCV, an active alarm<br />
indicates that delivered volume<br />
is less than 3 ml for three out <strong>of</strong><br />
four consecutive breaths.)<br />
Introduction<br />
0 to 99 L<br />
Accuracy:<br />
± (10 ml + 10% <strong>of</strong> reading)<br />
3 to 100 /minute<br />
Accuracy:<br />
± (0.1 +1% <strong>of</strong> setting)/minute<br />
20 to 6000 ml<br />
Accuracy:<br />
± (10 ml + 10% <strong>of</strong> setting)<br />
3 to 60 cmH 2 O (3 to 60 hPa)<br />
Accuracy:<br />
± (1 + 3% <strong>of</strong> setting)<br />
0 to 2000 ml<br />
Accuracy:<br />
± (10 ml + 10% <strong>of</strong> setting)<br />
700 Series Ventilator Operator’s Manual 10066984 Rev. D 1�27