Volume (continued)<br />

SPONT<br />

MINUTE<br />

VOLUME<br />

(L)<br />

(760 only)<br />

Shows the patient’s measured<br />

expiratory minute volume for all<br />

spontaneous breaths, based on<br />

the previous 60 seconds or 8<br />

breaths (whichever interval is<br />

shorter). Updated at the<br />

beginning <strong>of</strong> each breath.<br />

The calculation is reset when<br />

ventilation starts, when apnea<br />

ventilation starts or autoresets,<br />

when you change the mode or<br />

breath type, and when you press<br />

the alarm reset key.<br />

Alarm settings<br />

HIGH RATE An active alarm indicates that<br />

measured respiratory rate is<br />

higher than the alarm setting.<br />

HIGH<br />

TIDAL<br />

VOLUME<br />

LOW INSP<br />

PRESSURE<br />

LOW<br />

TIDAL<br />

VOLUME<br />

<strong>Table</strong> 1-4: 700 Series Ventilator keyboards: PATIENT DATA<br />

Key/<br />

indicator<br />

Function Range<br />

An active alarm indicates that<br />

exhaled volume for three out <strong>of</strong><br />

four consecutive breaths was<br />

above the alarm setting.<br />

An active alarm indicates that<br />

monitored circuit pressure is<br />

below the alarm setting at the<br />

end <strong>of</strong> inspiration. Inactive in for<br />

any spontaneous breath.<br />

An active alarm indicates that<br />

delivered volume for three out<br />

<strong>of</strong> four consecutive breaths were<br />

below the alarm setting. (If this<br />

alarm is set to 0 ml and breath<br />

type is PCV, an active alarm<br />

indicates that delivered volume<br />

is less than 3 ml for three out <strong>of</strong><br />

four consecutive breaths.)<br />

Introduction<br />

0 to 99 L<br />

Accuracy:<br />

± (10 ml + 10% <strong>of</strong> reading)<br />

3 to 100 /minute<br />

Accuracy:<br />

± (0.1 +1% <strong>of</strong> setting)/minute<br />

20 to 6000 ml<br />

Accuracy:<br />

± (10 ml + 10% <strong>of</strong> setting)<br />

3 to 60 cmH 2 O (3 to 60 hPa)<br />

Accuracy:<br />

± (1 + 3% <strong>of</strong> setting)<br />

0 to 2000 ml<br />

Accuracy:<br />

± (10 ml + 10% <strong>of</strong> setting)<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D 1�27