A.2.7 Every 2 years: device checks<br />

Maintenance<br />

<strong>Covidien</strong> recommends that you perform these checks every 2 years:<br />

1. Inspect the ventilator exterior for mechanical damage or<br />

evidence that the device has been subjected to a mechanical<br />

shock (for example, dropped). Any apparent damage constitutes<br />

a failure <strong>of</strong> this check.<br />

2. Run a full performance verification to verify that the ventilator<br />

functions properly, as described in the 700 Series Ventilator<br />

System Service Manual.<br />

3. Verify that the audible and visual alarms are functioning<br />

properly, as described in Appendix E.<br />

4. Inspect the labels for legibility.<br />

Record this data in an equipment log. If the ventilator fails any <strong>of</strong><br />

checks 1 through 3, remove it from use. Have a suitably qualified<br />

person service the ventilator and perform a full performance<br />

verification, as described in the 700 Series Ventilator System Service<br />

Manual. If the ventilator fails check 4, replace the labels, referring to<br />

the Service Manual for ordering information.<br />

A.2.8 Storage<br />

If you are storing the ventilator for 10 days or less, place the ventilator<br />

in standby mode (Chapter 3 tells you how) and disconnect the oxygen<br />

supply if you do not intend to use the ventilator immediately.<br />

If you are storing the ventilator for 6 months or longer, <strong>Covidien</strong><br />

recommends that a qualified service technician remove the internal<br />

battery (according to instructions in the 700 Series Ventilator System<br />

Service Manual), and that the external battery be disconnected.<br />

NOTE:<br />

A LOSS OF POWER alarm will sound for at least 2 minutes after<br />

you turn <strong>of</strong>f the power switch when no batteries are connected.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D A-13