Alarm testing<br />

NOTE:<br />

Run this test only with a fully charged battery. (Determine charge<br />

<strong>of</strong> internal battery by checking INTERNAL BATTERY LEVEL display.)<br />

12. If battery backup is available, the ventilator annunciates a<br />

LOSS AC POWER alarm. (If less than 5 minutes <strong>of</strong> battery backup<br />

are available, the ventilator annunciates a LOSS OF POWER<br />

alarm.)<br />

13. Reconnect the power cord to ac power. The LOSS OF POWER or<br />

LOSS AC POWER alarm should autoreset.<br />

14. APNEA alarm test: Press the MENU key, turn the knob to display<br />

User Settings, then press ACCEPT. Turn the knob to display<br />

Apnea interval (Ta), then press ACCEPT. Turn the knob to select<br />

an apnea interval <strong>of</strong> 20 seconds.<br />

15. Select the following settings:<br />

Mode SPONT<br />

APNEA PARAMS:<br />

RESPIRATORY RATE 5/min<br />

TIDAL VOLUME 400 L<br />

PEAK FLOW 10 L/min<br />

SUPPORT PRESSURE 0 cmH2O (0 hPa)<br />

The ventilator annunciates an APNEA alarm 20 seconds after the<br />

last breath triggered.<br />

NOTE:<br />

To avoid triggering a breath during the selected apnea interval,<br />

do not touch the test lung.<br />

16. Squeeze the test lung twice to simulate two subsequent<br />

patient-initiated breaths. The APNEA alarm should autoreset.<br />

700 Series Ventilator Operator’s Manual 10066984 Rev. D E-3