Self tests (SST and EST)<br />

If the SST<br />

result is:<br />

<strong>Table</strong> 4-3: Overall SST results<br />

It means: Do this:<br />

SST passed All tests passed. Press ACCEPT to exit SST<br />

or press the alarm reset<br />

key to restart SST from<br />

the beginning.<br />

SST<br />

incomplete<br />

All tests performed were<br />

passed, but some were<br />

skipped. The skipped tests<br />

were passed on a previous<br />

run.<br />

SST fault One or more tests did not<br />

pass. These tests might not<br />

compromise the ventilator’s<br />

ability to ventilate safely,<br />

based on the operator’s<br />

evaluation <strong>of</strong> the situation.<br />

SST failed One or more critical<br />

problems were detected.<br />

Press the alarm reset key<br />

to restart SST from the<br />

beginning or press<br />

MANUAL INSP to override<br />

this result and exit SST.*<br />

Press the alarm reset key<br />

to restart SST from the<br />

beginning or press<br />

MANUAL INSP to override<br />

this result and exit SST.*<br />

Press the alarm reset key to<br />

rerun SST. It is not possible<br />

to exit SST if SST fails. If SST<br />

continues to fail, remove<br />

the ventilator from use and<br />

contact service.<br />

* If you press MANUAL INSP to exit SST, you will be prompted to confirm that you<br />

wish to override the fault or incomplete test. Confirm by pressing MANUAL INSP<br />

again.<br />

4-12 10066984 Rev. D 700 Series Ventilator Operator’s Manual