Once ventilation begins<br />
4. Press MANUAL INSP; you do not have to hold the key down.<br />
The key lights and the ventilator displays this message:<br />
Man insp in progress<br />
5. When the manual inspiration is complete, the MANUAL INSP<br />
key light turns <strong>of</strong>f and breath delivery continues according to<br />
existing settings.<br />
5.6 The EXP PAUSE and INSP PAUSE keys (760 only)<br />
You can use the EXP PAUSE key to calculate<br />
the patient’s auto-PEEP (an estimate <strong>of</strong> the<br />
pressure that remains in the patient’s lungs<br />
at the end <strong>of</strong> exhalation, also known as<br />
intrinsic PEEP or occult PEEP). The presence<br />
<strong>of</strong> end expiratory flow (displayed as<br />
E: End xx in the message window) indicates<br />
the presence <strong>of</strong> auto-PEEP.<br />
EXP<br />
PAUSE<br />
7-00086<br />
1. Press the EXP PAUSE key during inspiration or exhalation: The<br />
key lights, and the ventilator closes the exhalation valve at the<br />
end <strong>of</strong> the expiratory phase without delivering a breath.<br />
Until the maneuver begins, the ventilator displays this message:<br />
Exp pause requested<br />
During the maneuver the ventilator displays this message:<br />
Exp pause active<br />
2. The EXP PAUSE maneuver ends when you release the key: hold<br />
the key down only until expiratory pressure stabilizes (watch<br />
the pressure bar graph or the message window).<br />
3. Once you release the EXP PAUSE key, the message window<br />
continues to display the patient’s calculated auto-PEEP for<br />
30 seconds:<br />
PEEP: i=xxx t=xxx<br />
where i is intrinsic PEEP and t is the total PEEP<br />
(total PEEP = measured PEEP + intrinsic PEEP).<br />
Once the maneuver is complete, the ventilator immediately<br />
delivers the mandatory breath that was suspended for the<br />
maneuver.<br />
5-10 10066984 Rev. D 700 Series Ventilator Operator’s Manual