<str<strong>on</strong>g>Tecnai</str<strong>on</strong>g> <strong>on</strong>-<strong>line</strong> <strong>help</strong> Opti<strong>on</strong>s 127Opti<strong>on</strong>s Software versi<strong>on</strong> 29.2.1 Tabular format reportThe tabular format list the camera and microscope settings at the top. Below that follows a table with thefollowing columns:• TEM - the magnificati<strong>on</strong> reported by the <str<strong>on</strong>g>Tecnai</str<strong>on</strong>g> microscope (the reference magnificati<strong>on</strong> <strong>on</strong> the platecamera, displayed by the microscope when the viewing screen is up).• X-grating - the results of the magnificati<strong>on</strong> calibrati<strong>on</strong> from the cross-grating directly.• Im. shift - the result of the magnificati<strong>on</strong> calibrati<strong>on</strong> performed using the image shift of themicroscope.• Lattice - the result of the magnificati<strong>on</strong> calibrati<strong>on</strong> performed with high-resoluti<strong>on</strong> images of gold [001]or silic<strong>on</strong> [110].