USER'S MANUAL<br />
General instructions for use<br />
6.7.4.2 <strong>Manual</strong> exposure<br />
In case the pre-set values are judged not sufficient for exam/patient, it is<br />
always possible to set values using the manual exposure mode.<br />
<br />
NOTE:<br />
In manual mode, take care that exposure parameters has to be set<br />
considering the type of images they are referring to; in fact, for<br />
longitudinal (or lateral) images kV value is increased passing from<br />
posterior to anterior teeth, and must be higher for scissors/canine to<br />
compensate the high X-ray absorption due to spine column. Vice versa,<br />
for cross images the value has to be lower for scissors/canine and higher<br />
for molar teeth.<br />
To modify kV or mA, the button "7" , or button "5" has to<br />
be pressed, respectively; in this way, the LED of button Patient size will<br />
be switched off, while LEDs of buttons Adult/Child and Implant will<br />
blink; display will show, per example, one of the following messages:<br />
or<br />
→ ← L O N G I T : 6 m m<br />
> 5 8 k V 0 7 m A 0 3 . 5 s<br />
→ ← T R A N S V : 4 m m<br />
5 8 k V > 0 7 m A 0 3 . 5 s<br />
while the symbol ">" will signal which parameter is to be modified.<br />
Using buttons Increase "3" and Decrease "6" , it will be<br />
possible to change the value of the selected parameter.<br />
kV value can be changed in the range 50÷80 kV, step 2 kV.<br />
mA value can be changed in the range 4÷10 mA, step 1 mA.<br />
<br />
NOTE:<br />
For a quick change, it is possible to speed up the process holding down<br />
the Increase button "3" or Decrease "6" .<br />
Select the Dental arch using the button Arch select large – medium (or<br />
standard) – narrow .<br />
(Rev. 1)<br />
93<br />
STRATO X (120V)