USER'S MANUAL<br />

General instructions for use<br />

6.4.2.2 <strong>Manual</strong> exposure<br />

If the kV and mA combinations of the above table are not considered<br />

suitable for a specific examination, it will be possible to set new<br />

parameters using the manual mode.<br />

To change the kV or mA values, press key "7" , or key "5" .<br />

The LED on the Patient Size key switches off and the Adult / Child and<br />

Panoramic keys will flash. One of the two following messages will be<br />

displayed:<br />

or<br />

> 7 4 k V 1 0 m A 1 5 . 0 s<br />

7 4 k V > 1 0 m A 1 5 . 0 s<br />

The symbol ">" indicates which parameter is being changed.<br />

The selected parameter can be changed using the increase key "3"<br />

and decrease key "6" .<br />

The kV value can vary between 50 and 80 kV, with 2 kV steps.<br />

The mA value can vary between 4 and 10 mA, with 1 mA steps.<br />

<br />

NOTE:<br />

To change the values rapidly, keep the increase key "3"<br />

decrease key "6"<br />

pressed.<br />

or<br />

Select the mouth structure using the Arch wide - medium - narrow<br />

key .<br />

STRATO X (120V) 42<br />

(Rev. 1)