USER'S MANUAL<br />
General instructions for use<br />
6.6.2.2 <strong>Manual</strong> exposure<br />
If the kV and mA combinations of the above table are not considered<br />
suitable for a specific examination, it will be possible to set new<br />
parameters using the manual mode.<br />
To change the kV or mA values, press key "7" , or key<br />
"5" . The LED on the Patient Size key switches off and the<br />
Adult / Child and Biaxial TMJ keys will flash. One of the two following<br />
messages will be displayed:<br />
or<br />
> 7 4 k V 1 0 m A 1 0 . 5 s<br />
7 4 k V > 1 0 m A 1 0 . 5 s<br />
The symbol ">" indicates which parameter is being changed.<br />
The selected parameter can be changed using the increase key "3"<br />
and decrease key "6" .<br />
The kV value can vary between 50 and 80 kV, with 2 kV steps.<br />
The mA value can vary between 4 and 10 mA, with 1 mA steps.<br />
<br />
NOTE:<br />
To change the values rapidly, keep the increase key "3"<br />
decrease key "6"<br />
pressed.<br />
or<br />
(Rev. 1)<br />
69<br />
STRATO X (120V)