USER'S MANUAL<br />

Safety information<br />

3. CLEANING AND DISINFECTION<br />

In order to get an accurate hygene and cleaning it is necessary to follow<br />

thoroughly all indicated procedures.<br />

<br />

WARNING:<br />

Disconnect the unit from the mains<br />

before performing any cleaning.<br />

Do not let water or liquids enter the unit, these could cause corrosion or<br />

short-circuits.<br />

Use only a wet cloth and a mild<br />

detergent to clean the painted<br />

surfaces, the accessories and the<br />

connection cables, and wipe then<br />

with a dry cloth; do not use<br />

solvents (alcohol, benzine, trichloroethylene)<br />

nor corrosive and abrasive<br />

solutions.<br />

To clean the rare earths scintillating screens in the cassette, please follow<br />

the indication given by its manufacturer, do not use, detergents, solvent<br />

(alcohol, benzine), corrosive or abrasive stuffs.<br />

The bite-holding rod, the centring bite and the ear centring units of<br />

the cephalostatus must be replaced after each examination.<br />

Clean thoroughly the chin rest, the handles, the nose-rest and the<br />

temple-support any time these are used.<br />

It is advisable to disinfect, whenever necessary, the chin-rest, the<br />

handles, the nose-rest and the temple support with a 2% Glutaraldehyde<br />

solution or similar (find out what is available; e.g. Milton).<br />

(Rev. 1)<br />

7<br />

STRATO X (120V)