USER'S MANUAL<br />

General instructions for use<br />

5. The cassette can now be opened in a dark room and the film<br />

developed.<br />

<br />

NOTE:<br />

After every examination, clean the ear centering device and temple<br />

support thoroughly.<br />

<br />

NOTE:<br />

If, during the exposure the patient moves or the operator realizes that the<br />

parameters set are incorrect, the X-ray button "30" must be released<br />

immediately, thus suspending emission of the X-rays.<br />

The following message will be displayed:<br />

E 5 6 1<br />

P R E S S R E S E T<br />

Press button "9"<br />

. The following message will be displayed:<br />

C E P H<br />

R E M O V E F I L M<br />

The system now returns to the position for Ceph exam and the unit<br />

starts the procedure for the new examination.<br />

<br />

NOTE:<br />

We recommend verifying that the system is still aligned correctly with the<br />

patient (Frankfurt plane) via the laser centering devices.<br />

<br />

NOTE:<br />

No warning message will be given if the cassette is removed but the film<br />

is not replaced, and another exposure is made. However, the film would<br />

provide non-diagnostic results due to the double exposure.<br />

(Rev. 1)<br />