USER'S MANUAL<br />

Technical specifications<br />

5.5 Measurement method of technical factors<br />

(Paragraph for the authorized personnel)<br />

<br />

WARNING:<br />

These measurements require the removal of the HF group covers; this<br />

means to gain access to internal parts where high voltage are normally<br />

present.<br />

For the measurement of the exposure parameters with the invasive<br />

method, please follow the procedure described in paragraphs 7.2 and 7.3<br />

of the Service manual.<br />

<br />

WARNING:<br />

During the panoramic examination, the set value of kV and tube current<br />

varies according to a pre-determined curve in order to compensate the<br />

different absorption of X-ray beam due to different anatomical structures.<br />

In this way, it is possible to obtain a good uniformity of the image’s<br />

contrast. Particularly, the chosen value is lowered on the initial phase of<br />

the panoramic and increased on the scissors/canine zone, in order to<br />

compensate the effect of the cervical spine.<br />

The value displayed during the panoramic examination corresponds to<br />

the to chosen one, while the instantaneous value can be different; these<br />

effects must be considered in case of measure of the exposure factors<br />

using standard diagnostic mode (please refer to the note at page 51).<br />

Accuracy declared on the section "Technical data" is referred to the<br />

actual value of kV and/or mA. In any case, manufacturer guarantees<br />

that the accuracy of loading factors is always compliance with the<br />

international standard for safety of medical devices IEC 601-1.<br />

Particularly, in accordance with IEC 601-2-7, the maximum deviation<br />

(including the correction and instrument’s accuracy) is less than or equal<br />

to ±10 for kV, while for tube current is less than or equal to ±15%.<br />

STRATO X (120V) 26<br />

(Rev. 1)