USER'S MANUAL<br />

General instructions for use<br />

• The image shows TMJ at different height<br />

This fact is described in Figure 44.<br />

Figure 44<br />

Possible cause:<br />

The mid sagittal plane is not vertical. This can be a patient’s<br />

problem, but if the defect is always present, check the laser beam.<br />

Remedy:<br />

Verify that the laser beam is vertical; this check can be performed<br />

very quicly using the laser beam and verifying that falls on the center<br />

of the chin support; remove the chin support itself and check that<br />

the beam falls on the center of the holes used to fix the chin itself.<br />

If not, a possible cause can be due to a non perfect horizontally of the<br />

chin arm.<br />

STRATO X (120V) 128<br />

(Rev. 1)